WRIGHTS LANE SYNTHES USA PRODUCTS LLC HOLDING SLEEVE 105MM; GUIDE, SURGICAL, INSTRUMENT
|
Back to Search Results |
|
| Model Number |
394.46 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2022, patient underwent an unknown surgical procedure for unknown reason.While working with a large distractor, the wing nut screw broke off inside of the holding sleeve after tightening the schanz screw.The schanz screw was removed with the attached holding sleeve and the procedure was successfully completed.There was no surgical delay reported.There were fragments generated and removed without additional intervention.There was no patient consequence reported.This report is for one (1) holding sleeve 105mm this is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d4 d10: date of concomitant therapy is january 6, 2022.H3, h6: a manufacturing related potential cause was not suspected, therefore, per franchise complaint product investigation procedure, no manufacturing record evaluation is required.Visual inspection: the holdsleeve w/wingscr l105 i-ø 6 f/large- was received at juarez pal.Visual examination of the returned device found the wing screw has broken off.No other product defects were found.Dimensional inspection: no dimensional inspection was performed due to post-manufacturing damage.Complaint relevant dimensions cannot be taken.Document/specification review: the current revision of drawings were reviewed.No design issues or discrepancies were identified.Investigation conclusion: the overall complaint was confirmed for the received device.No definitive root cause could be determined based on the provided information.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D9.
|
| |
|
Search Alerts/Recalls
|
|
|