• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. AEROAUTOCAT 2 WAVE; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. AEROAUTOCAT 2 WAVE; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN000329
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/05/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the pump shut down while running on battery power.The pump alarmed multiple times relating to helium loss.The pump was swapped at the destination.There was no report of patient complications, serious injury or death.
 
Additional Manufacturer Narrative
(b)(4).
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that the pump shut down while running on battery power.The pump alarmed multiple times relating to helium loss.The pump was swapped at the destination.There was no report of patient complications, serious injury or death.
 
Additional Manufacturer Narrative
Qn# (b)(4).Teleflex received the device for investigation.The reported complaint of "pump shut down running on battery power" is confirmed.The pump shut off after approximately 72 minutes when operating on battery power after a full charge during the complaint investigation.The battery failed battery load test.A definitive root cause could not be determined but a potential cause of the short battery life is a result of battery maintenance.Iabp users and servicers must follow the operating instructions manual for recommendations on usage, charging, maintenance and storage of the batteries.If battery maintenance is not performed per the iabp operating instructions manual, the battery may not provide the expected minimum run time of operating power.A device history record (dhr) review was conducted for the serial number/lot number with no relevant findings.The device passed all manufacturing specifications prior to release.Teleflex assessed the risk for the reported complaint.There are no new or revised risk.This will be monitored for any developing trends.
 
Event or Problem Description
It was reported that the pump shut down while running on battery power.The pump alarmed multiple times relating to helium loss.The pump was swapped at the destination.There was no report of patient complications, serious injury or death.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AEROAUTOCAT 2 WAVE
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
kaylia pen
16 elizabeth drive
chelmsford, MA 01824
MDR Report Key13451122
Report Number3010532612-2022-00033
Device Sequence Number4675213
Product Code DSP
UDI-Device Identifier30801902051739
UDI-Public30801902051739
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K060309
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 01/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN000329
Device Catalogue NumberIAP-0535
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/05/2022
Supplement Date Received by Manufacturer03/04/2022
Initial Report FDA Received Date02/04/2022
Supplement Report FDA Received Date03/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-