SMITH & NEPHEW, INC. G2 CONST INSRT SZ1-2 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 71420960
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fever (1858); Unspecified Infection (1930); Swelling/ Edema (4577)
Date of Event 01/14/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: case (b)(4).
 
Event or Problem Description
It was reported that after a tka performed on (b)(6) 2020, the patient experienced swelling, fever and infection.This adverse event was addressed via revision surgery on (b)(6) 2022.The g2 const insrt sz1-2 11mm was explanted.The doctor said the devices were not defective.Patient's current status is unknown.
 
Additional Manufacturer Narrative
H3,h6: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, per complaint details, a revision was performed approximately 22 months post implantation due to swelling, fever and infection.Reportedly, the doctor said the devices had ¿no problem¿.Per correspondence, the requested documentation and current patient status are not available.Reportedly, the infection was the root cause of the revision; however, the source of the infection remains unknown.Without the requested documentation, no further root cause analysis can be performed and the patient impact beyond that which was reported cannot be determined.No further medical assessment can be rendered at this time.A review of complaint history for the previous 12 months did not reveal similar events for the listed device.A review of the instructions for use documents for knee system revealed that wound infection has been identified as a possible adverse effects.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but are not limited to contamination, patient reaction, and post-operative healing issue.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G2 CONST INSRT SZ1-2 11MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13451209
Report Number1020279-2022-00489
Device Sequence Number12235604
Product Code JWH
UDI-Device Identifier03596010210104
UDI-Public03596010210104
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K962137
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71420960
Device Catalogue Number71420960
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/14/2022
Supplement Date Received by Manufacturer03/16/2022
Initial Report FDA Received Date02/04/2022
Supplement Report FDA Received Date03/18/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-