MEDTRONIC, INC. RESTORE; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS Back to Search Results
Model Number 37022
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/03/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Other relevant device(s) are: product id: neu_unknown_lead, serial/lot #: unknown.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during stage 1 lead implant, reported one side tested good, buried lead and went to the other side implanted lead and found segment 2a and 2b >30,000 ohms.An attempt was made to resolve the issue, by replacing tablet, ens, cable and then lead.Caller reached out to technical services after lead replacement, and now 2b and 2c are >30 ,000 ohms.It was decided to implant lead and check impedances when ins is connected.Caller needed to disconnect call - will email rep to get required information.
 
Event or Problem Description
Additional information was received from a manufacturer representative (rep) reporting that they were getting configuration not verified but this was resolved by re-clipping and positioning the test cable.The rep tested at home with a new lead and ens submerged in water and still replicated the high impedances on contact 2c (>10k ohms) even after letting voltage ramp up with impedance check.The rep was using a sensight lead and had swapped to a different tablet owned by the hcp.They added that a colleague and hcp had seen this before but appears to resolve itself after stage 2.The rep indicated that the cause of the high impedance was not determined.The high impedances on contact 2 persisted.
 
Additional Manufacturer Narrative
Continuation of d10: product id b3301542m serial# (b)(6) product type lead.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
D10: section d information references the main component of the system.Other relevant device(s) are: product id b3301542m serial# (b)(6) product type lead product id 37022 lot# serial# (b)(6) product type external neurostimulator product id b3301542m lot# serial# (b)(6) product type lead medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information was received from a manufacturer representative (rep) reporting that two enss were used.
 
Event or Problem Description
Additional information was received the manufacturer representative: high impedances resolved with stage 2 procedure.
 
Additional Manufacturer Narrative
Continuation of d10: product id b3301542m, serial# (b)(6), implanted: (b)(6) 2022, explanted: (b)(6) 2022, product type lead product id 37022, serial# (b)(6), product type external neurostimulator product id b3301542m, serial# (b)(6), implanted: (b)(6) 2022, product type lead.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Continuation of d10: product id b3301542m lot# serial# (b)(6) implanted: 2022-(b)(6) explanted: 2022-(b)(6) product type lead product id 37022 lot# serial# (b)(6) implanted: explanted: product type external neurostimulator product id b3301542m lot# serial# (b)(6) implanted: 2022-(b)(6) explanted: product type lead product id neu_stylet_acc lot# serial# unknown implanted: explanted: product type accessory h3: analysis of the lead va2grcrv09 found no anomalies.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESTORE
Common Device Name
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key13453102
Report Number2182207-2022-00189
Device Sequence Number14925774
Product Code MHY
UDI-Device Identifier00643169513013
UDI-Public00643169513013
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P960009
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 04/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number37022
Device Catalogue Number37022
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/03/2022
Supplement Date Received by Manufacturer02/07/2022
02/16/2022
02/25/2022
04/01/2022
Initial Report FDA Received Date02/04/2022
Supplement Report FDA Received Date02/11/2022
02/21/2022
02/28/2022
04/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/28/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"SEE H10...."; SEE H10
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient SexFemale
-
-