• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. OPTETRAK; LOGIC TIBIA PS MOD INSRT SZ 5 13MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. OPTETRAK; LOGIC TIBIA PS MOD INSRT SZ 5 13MM Back to Search Results
Model Number 02-012-35-5013
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 06/17/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Pending evaluation.
 
Event or Problem Description
As reported, approximately 3.86 years since initial implant, this male patient (b)(6) underwent a left side tka and was implanted with an exactech uhmwpe insert, serial # (b)(4).On (b)(6) 2021, the patient underwent revision due to aseptic loosening.No additional information is available or forthcoming.
 
Additional Manufacturer Narrative
Pending evaluation.
 
Event or Problem Description
As reported, approximately 3.86 years since initial implant, this male patient (b)(6) underwent a left side tka and was implanted with an exactech uhmwpe insert, serial # (b)(4).On (b)(6) 2021, the patient underwent revision due to aseptic loosening.No additional information is available or forthcoming.
 
Additional Manufacturer Narrative
Section h10: (h3) the revision reported was likely the result of an insufficient bond between the implant and the bone and patient-related conditions, which led to aseptic (non-infected) loosening.However, this cannot be confirmed as the devices were not available for evaluation and pre-revision x-rays were not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTETRAK
Common Device Name
LOGIC TIBIA PS MOD INSRT SZ 5 13MM
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 ct
gainesville FL 32653
Manufacturer Contact
kate jacobson
3523771140
MDR Report Key13453326
Report Number1038671-2022-00177
Device Sequence Number5046842
Product Code JWH
UDI-Device Identifier10885862001948
UDI-Public10885862001948
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K033883
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/18/2019
Device Model Number02-012-35-5013
Device Catalogue Number02-012-35-5013
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/07/2022
Supplement Date Received by Manufacturer06/09/2022
Initial Report FDA Received Date02/04/2022
Supplement Report FDA Received Date07/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/29/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
-
-