DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Incomplete or Inadequate Connection (4037)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/10/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Testing of actual/suspected device|10: a getinge field service engineer (fse) evaluated the cardiosave intra-aortic balloon pump (iabp) unit and was able to reproduce the reported issue.The fse performed a fiber optic test which failed.The fse noted that the fiber optic waveform was working but no transducer waveform was observed.To fix the issue, the fse replaced the cbl assy,fo sensor extension and performed all functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that prior to use, the cardiosave intra- aortic balloon pump (iabp) unit had a fiber optic connection failure.There was no patient involvement, and no adverse event reported.
|
|
Event Description
|
N/a.
|
|