A visual, dimensional, and functional analysis was performed on the returned device.The reported difficulty to insert was unable to be confirmed due to the condition of the returned dc pod.The damage to the delivery catheter pod was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history revealed no other similar complaints reported from this lot.The investigation was unable to determine a conclusive cause for the reported difficulties.It may be possible that the filter was pulled into the pod too quickly or with excessive force causing damage to the delivery catheter pod preventing the filtration element from being able to load properly; however, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling.
|
It was reported that when during preparation of the emboshield nav6 embolic protection system (eps), the filter could not be loaded into the delivery catheter pod and the tip of the delivery catheter was noted to be kinked.There was bunching/wrinkling noted on the delivery catheter pod but it was observed after loading was attempted.Therefore, the device was not used and there was no patient involvement.Another emboshield nav6 was used to complete the procedure.There was no clinically significant delay in the procedure.Returned device analysis identified that the dc pod was separated and torn and the separated portion was not returned.Follow up with the account confirmed they were not aware and the separated portion was discarded.No additional information was provided.
|