BOSTON SCIENTIFIC CORPORATION NC QUANTUM APEX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 6534
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in a moderately calcified vessel.A 30mm x 3.50mm nc quantum apex balloon catheter was advanced for dilatation.However, during inflation at 10 atmospheres, the balloon broke.The procedure was completed with a different device.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of an nc quantum apex balloon catheter.The device was visually and microscopically examined.There were numerous kinks to the hypotube of the device.There was contrast in the inflation lumen and balloon.There was blood in the guidewire lumen, and the balloon was loosely folded.At 4mm from the tip moving proximally for a length of 18mm, the balloon has a longitudinal tear.Product analysis confirmed the reported burst, as there was a longitudinal tear in the balloon.
 
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in a moderately calcified vessel.A 30mm x 3.50mm nc quantum apex balloon catheter was advanced for dilatation.However, during inflation at 10 atmospheres, the balloon broke.The procedure was completed with a different device.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC QUANTUM APEX
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13455519
Report Number2134265-2022-01205
Device Sequence Number6009572
Product Code LOX
UDI-Device Identifier08714729783619
UDI-Public08714729783619
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K121667
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/04/2024
Device Model Number6534
Device Catalogue Number6534
Device Lot Number0028141878
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/01/2022
Supplement Date Received by Manufacturer05/23/2022
Initial Report FDA Received Date02/04/2022
Supplement Report FDA Received Date05/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/05/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-