Model Number 367963
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported when using the bd vacutainer® pst¿ gel and lithium heparinn (lh) blood collection tube there was a broken lid/cap/hemogard shield.The following information was provided by the initial reporter.The customer stated: "the rubber stopper was assembled incorrectly.".
 
Additional Manufacturer Narrative
H.6.Investigation: bd received 2 photos from the customer in support of this complaint.A visual examination of photos was performed and revealed improper assembly.The improper assembly of the shield and stopper can happen at the metro where the stopper is placed/pushed into the shield.If a stopper goes into the rail crooked or cocked it can potentially go into a shield and be placed on a tube.Bd was able to confirm the customer¿s indicated failure mode with the photos provided.The device history records were reviewed with no issues being identified.There were no related quality notifications.All process and final inspections comply with specification requirements h3 other text : see h.10.
 
Event or Problem Description
It was reported when using the bd vacutainer® pst¿ gel and lithium heparin (lh) blood collection tube there was a broken lid/cap/hemogard shield.The following information was provided by the initial reporter.The customer stated: "the rubber stopper was assembled incorrectly.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBE
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13455637
Report Number1917413-2022-00050
Device Sequence Number4853827
Product Code JKA
UDI-Device Identifier50382903679633
UDI-Public50382903679633
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K945952
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date08/31/2022
Device Model Number367963
Device Catalogue Number367963
Device Lot Number1228175
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/26/2022
Supplement Date Received by Manufacturer02/25/2022
Initial Report FDA Received Date02/04/2022
Supplement Report FDA Received Date03/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-