Model Number AED PLUS
Medical Device Problem Code Unable to Obtain Readings (1516)
Health Effect - Clinical Code Necrosis (1971)
Date of Event 01/17/2022
Type of Reportable Event Death
Event or Problem Description
Complainant alleged that while attempting to treat a patient (age & gender unknown), the device displayed a "check pads" message.Complainant indicated that the patient subsequently expired.
 
Additional Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Additional Manufacturer Narrative
The device was returned to zoll medical canada and there was no indication of a device malfunction.The device passed all testing using the returned pads and a known good set of pads without duplicating the report.The device was recertified and returned to the customer.Review of the device logs shows evidence of poor coupling.There are several padz off messages indicating that the patient impedance had gone out of range.The ecg waveform acquired has artifacts, baseline wandering, and signal railing.Electrode labeling states the importance of good placement on the patient and provides instruction for proper electrode application technique.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AED PLUS
Common Device Name
DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key13456733
Report Number1220908-2022-00417
Device Sequence Number12296526
Product Code MKJ
UDI-Device Identifier00847946024222
UDI-Public00847946024222
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K011541/P160
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Model NumberAED PLUS
Device Catalogue NumberAED PLUS
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/25/2022
Supplement Date Received by Manufacturer01/25/2022
Initial Report FDA Received Date02/04/2022
Supplement Report FDA Received Date03/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/01/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient SexUnknown
-
-