| Catalog Number |
195-160 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The remainder of the investigation remains in progress.A supplemental report will be provided after completion.
|
| |
|
Event or Problem Description
|
|
The consumer reported conflicting results with two binaxnow¿ covid-19 antigen self tests performed on (b)(6) 2022.The result from the first test was negative.The repeat test generated a positive result.No additional patient information, including treatment and outcome, was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The remainder of the investigation remains in progress.A supplemental report will be provided after completion.
|
| |
|
Event or Problem Description
|
|
The consumer reported conflicting results with two binaxnow¿ covid-19 antigen self tests performed on (b)(6) 2022.The result from the first test was negative.The repeat test generated a positive result.No additional patient information, including treatment and outcome, was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation report: testing was performed in triplicate at abbott diagnostics scarborough, inc.On retained kit lot 177486 with internal positive quality control samples and negative quality control swabs.All test results were valid and performed as expected.Additionally, the manufacturing records and quality control release testing was reviewed for test kit part number 195-160/ lot 177486 and device part number 195-430h / lot 174663.The lot met the required release specifications.A review of the complaints reported as false negative patient results (confirmed and unconfirmed, conflicting results) related to kit lot 177486 showed that the complaint rate is (b)(4).A review of the complaints reported as false positive patient results (confirmed and unconfirmed, conflicting results) related to kit lot 177486 showed that the complaint rate is (b)(4).Abbott diagnostics scarborough, inc.Was unable to determine the exact root cause of the reported issue.However, it could possibly be related to issues including the self-test user performance, interpretation of the result, or the specific patient sample.
|
| |
|