| Model Number |
0684-00-0479-01C |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that after 2 days of intra-aortic balloon(iab) therapy, blood was seen in the tubing.The iab was removed and lumen was found to be broken.There was no reported injury or adverse event to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated fields: b4, d10, g4, h3, h6, h10.The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.Complaint record id (b)(4).
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned with the membrane completely unfolded with blood on the interior and exterior of the catheter and between the catheter and the sheath.The returned sheath was not a maquet product.The extender tubing was also returned.A visual examination of the product detected the inner lumen within the membrane was completely separated within a kink approximately 24.6cm from iab tip.The inner lumen was found occluded with dried blood.The occlusion was unable to be cleared.An underwater leak test of the balloon, catheter, y-fitting, extracorporeal and extender tubing was performed and no other leaks were detected.The break found in the inner lumen appears to have been the result of a severe kink, which eventually failed causing reported problems.It is difficult to determine when the break occurred but it is possibly a result of patient movement during the procedure.A non-conforming material report review was completed for the reported product.No nonconformances were found that are considered to be related to the event.Reference complaint#: (b)(4).
|
| |
|