Catalog Number 383012 |
Device Problem
Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/17/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that rust was found on the bd intima ii¿ iv catheter prn adapter needle.The following information was provided by the initial reporter, translated from (b)(6): "needle rust was found when indwelling needle was used for patient".
|
|
Manufacturer Narrative
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Manufacturer Narrative
|
H6: investigation summary: a device history review was conducted for lot number 1076999.Our records show that this is the only instance of this issue occurring in this production batch.According to the sampling plan applied for product performance, this lot was accepted and released without defects being noted during the final assembly or visual inspections.A sample could not be obtained for evaluation and testing; in lieu of the affected device, visual analysis was performed on retention samples for this lot, the retained devices were found to be free of any damage or other abnormalities.Unfortunately without the ability to investigate the affected unit our quality engineers were unable to determine the root cause for this complaint.
|
|
Event Description
|
It was reported that rust was found on the bd intima ii¿ iv catheter prn adapter needle.The following information was provided by the initial reporter, translated from chinese: "needle rust was found when indwelling needle was used for patient".
|
|