• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FLEXIVA; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FLEXIVA; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number M0068403920
Medical Device Problem Codes Degraded (1153); Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/10/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2022 that a flexiva laser fiber was used in a ureteroscopy stone removal procedure in the distal ureter performed on (b)(6) 2022.The laser unit used was a lumenis 100 laser console with laser settings of 0.6 joules, rate 50.During the procedure, after less than 10 minutes, the laser beam was lost and the connector of the laser fiber was super hot to touch.The procedure was completed with another flexiva laser fiber.There were no patient or user complications reported as a result of this event.
 
Additional Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIVA
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
COHERENT INCORPORATED
5100 patrick henry drive
santa clara CA 95054
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13457656
Report Number3005099803-2022-00488
Device Sequence Number20383296
Product Code GEX
UDI-Device Identifier08714729808183
UDI-Public08714729808183
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K100078
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/15/2024
Device Model NumberM0068403920
Device Catalogue Number840-392
Device Lot Number0000011355
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/11/2022
Initial Report FDA Received Date02/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age33 YR
Patient SexMale
Patient Weight85 KG
-
-