MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. VERSAJET II CONSOLE; LAVAGE, JET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. VERSAJET II CONSOLE; LAVAGE, JET Back to Search Results
Model Number 66800039
Medical Device Problem Codes Failure to Prime (1492); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/18/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, during debridement surgery, water failed to reach the handpiece of a versajet ii hydrosurgery system console.By the time this happened, the patient was already anaesthetized, and it was reported that a lot of time was lost in the operating room.The console started up properly.The purge was done, but there was no water in the handpiece.The procedure could not be successfully completed, for which it was cancelled and postponed.No injury caused to the patient.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, during debridement surgery, water failed to reach the handpiece of a versajet ii hydrosurgery system console.By the time this happened, the patient was already anaesthetized, and it was reported that a lot of time was lost in the operating room.The console started up properly.The purge was done, but there was no water in the handpiece.The procedure could not be successfully completed, for which it was cancelled and postponed.No injury caused to the patient.
 
Additional Manufacturer Narrative
The investigation is now complete.The device was returned for evaluation and with no faults found, the event cannot be confirmed.A visual inspection observed signs of general wear and tear, the functional evaluation found no device faults.Utilizing a test handpiece, the device primed and operated within expected parameters.A documentation review has been conducted, confirming previous complaints of this nature, the device history and historical review have observed no manufacturing quality concerns.The risk files mitigate the reported event which is further delineated in the instruction for use (ifu).The root cause could not be determined as no fault was found.The ifu explicitly states to ensure that the device is fully primed and operational prior to the administration of anesthesia, to prevent unnecessary delays.No manufacturing related concerns have been identified, therefore no corrective action is deemed necessary.Smith and nephew can confirm the device was released according to specifications and continue to monitor for adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAJET II CONSOLE
Common Device Name
LAVAGE, JET
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13457972
Report Number8043484-2022-00049
Device Sequence Number19990532
Product Code FQH
UDI-Device Identifier00040565124759
UDI-Public00040565124759
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K143115
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number66800039
Device Catalogue Number66800039
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/18/2022
Supplement Date Received by Manufacturer03/26/2022
Initial Report FDA Received Date02/04/2022
Supplement Report FDA Received Date04/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-