MAUDE Adverse Event Report: ABBOTT VASCULAR HI-TORQUE COMMAND 18 GUIDE WIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR HI-TORQUE COMMAND 18 GUIDE WIRE Back to Search Results
Model Number 1013753
Medical Device Problem Codes Peeled/Delaminated (1454); Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was received.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported the procedure was performed to treat a lesion in the left anterior tibial artery with moderate calcification and mild tortuosity.A command 18 guide wire was advanced in the vessel without resistance, followed by an armada 18 balloon catheter.When the command guide wire was being removed there was resistance with the balloon catheter.Both devices had to be removed as one unit.Once outside the patient anatomy, it was noted that coating of the guide wire had peeled off the tip of the guide wire.The armada 18 balloon catheter was flushed, and the separated coating was flushed completely.There was no coating in the patient's anatomy present.A steelcore guide wire was used with the same armada 18 balloon catheter without issue.There was no issue with the armada 18 balloon catheter.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
A visual inspection was performed on the returned guide wire and the reported peeling was confirmed.The reported difficulty to remove from the balloon catheter was unable to be confirmed due to the returned condition of the guide wire.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents and/or complaint reported from this lot.There was no damage noted to the guide wire during the inspection prior to use which suggests a product quality issue with respect to manufacture, design or labeling did not contribute to the reported difficulties.The investigation determined the reported difficulty to remove and polymer peeling appear to be related to case circumstances of the procedure.In this case, based on the reported information, it is likely that during use, the polymer became damaged against the anatomy and/or other devices used resulted in the reported difficulty to remove from the balloon catheter.Manipulation against resistance ultimately resulted in the polymer tears and separation causing the appearance of peeling.The noted bends on the guide wire likely occurred during packing for return analysis.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE COMMAND 18 GUIDE WIRE
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13458086
Report Number2024168-2022-01210
Device Sequence Number4675444
Product Code DQX
UDI-Device Identifier08717648211980
UDI-Public08717648211980
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K152404
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2022
Device Model Number1013753
Device Catalogue Number1013753
Device Lot Number0121561
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/26/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/18/2022
Supplement Date Received by Manufacturer02/21/2022
Initial Report FDA Received Date02/04/2022
Supplement Report FDA Received Date03/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/15/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
6FR SHEATH.; ARMADA 18 BALLOON CATHETER.
Patient SexUnknown
-
-