Model Number 9600TFX26A
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Rupture (2208)
Event Date 01/11/2022
Event Type  Injury  
Manufacturer Narrative
Per the instructions for use (ifu), cardiovascular injuries such as perforation or dissection of vessels, ventricle, myocardial or valvular structures including rupture of the rvot, and hematoma are known potential adverse events associated with the overall thv procedure and may require intervention.According to the thv training manuals, risk factors for aortic dissection, hematoma, annular rupture, rupture of rvot during the thv procedure include significant thv over sizing, severely obliterated sinuses of valsalva, porcelain aorta and/or presence of bulky calcification, and narrow calcified stj.In addition, advanced age, female gender, small body weight, and steroid dependency can also be contributing factors.The sapien valve relies on native valve calcium to securely anchor to the annulus.Despite this beneficial aspect of calcium, bulky calcium can increase the risk of calcific nodule displacement into the vasculature, which can lead to vascular injury.At times the extent and distribution of calcium can impair ease of delivery of the valve, correct positioning of the valve, deployment of the valve, and procedural success.Per the instructions for use (ifu), rupture of the rvot is a potential adverse event associated with the transcatheter valve replacement (tvr) procedure and the use of the edwards thv devices.In this case, there was no allegation or indication a product malfunction contributed to this adverse event.Investigation results suggest/indicate patient factors may have caused or contributed to this event.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required.
 
Event Description
As reported, during a transfemoral transcatheter pulmonic valve replacement (tpvr) with a 26mm sapien 3 valve in a previously implanted conduit, post implant, the valve was explanted due to a rupture of the pulmonary/rvot conduit and replaced with a surgical valve.The patient is stable post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS SAPIEN 3 TRANSCATHETER HEART VALVE
Type of Device
PULMONARY VALVE PROSTHESIS PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
renee van dorne
1 edwards way
irvine, CA 92614
9492506385
MDR Report Key13458549
MDR Text Key287421290
Report Number2015691-2022-03793
Device Sequence Number1
Product Code NPV
UDI-Device Identifier00690103194357
UDI-Public(01)00690103194357(17)230603
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P200015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/03/2023
Device Model Number9600TFX26A
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/11/2022
Initial Date FDA Received02/04/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/21/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexMale
-
-