| Model Number |
MICL13.2 |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
01/07/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 13.2mm micl13.2 implantable collamer lens, -11.0 diopter into the patient's left (os) eye on (b)(6) 2021.The surgeon reports vaulting and corneal touch.Attempts to obtain additional information have not been successful.
|
| |
|
Additional Manufacturer Narrative
|
|
Sex, weight, race, ethnicity: unk.Date of event: unk.(b)(4).Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
B5-additional information: excessive vaulting is reported.On (b)(6) 2022 the lens was explanted and replaced with a shorter lens and the problem is resolved.Mild pigment dispersion without symptoms after exchange is reported, which has resolved.The cause of the event is reported as a patient-related factor; the lens did not fail to perform as intended: "poor estimate of ciliary sulcus diameter relative to horizontal wtw." claim# (b)(4).
|
| |
|
Search Alerts/Recalls
|