Model Number G247
Device Problem Pocket Stimulation (1463)
Patient Problem Device Overstimulation of Tissue (1991)
Event Date 12/19/2021
Event Type  Injury  
Event Description
It was reported that the patient implanted with this cardiac resynchronization therapy defibrillator (crt-d) device felt little shocks in the diaphragm area.Boston scientific technical services (ts) discussed that possibly it is the left ventricular (lv) lead stimulating the diaphragm.Subsequently, this device was explanted.No additional adverse patient effects were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient implanted with this cardiac resynchronization therapy defibrillator (crt-d) device felt little shocks in the diaphragm area.Boston scientific technical services (ts) discussed that possibly it is the left ventricular (lv) lead stimulating the diaphragm.Subsequently, this device was explanted.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIGILANT X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key13462338
MDR Text Key285330494
Report Number2124215-2022-03422
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589287
UDI-Public00802526589287
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 02/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/04/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/10/2023
Device Model NumberG247
Device Catalogue NumberG247
Device Lot Number271817
Was Device Available for Evaluation? No
Date Manufacturer Received12/19/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/17/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age53 YR
Patient SexMale
-
-