BOSTON SCIENTIFIC CORPORATION ALTRUA 60; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
|
Back to Search Results |
|
| Model Number |
S603 |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/03/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this device showed half a year remaining from two months ago with magnet rate of ninety.Recently, device battery was at one year remaining with the same magnet rate.Boston scientific technical services (ts) discussed battery performance would vary a lot in this device in which it could be that the low estimate was right after a telemetry session that took longer or lead testing.Ts then advised to save to disk for analysis.To date, the device remains in service.No adverse patient effects reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that battery status indicators were different than expected based on information provided to boston scientific.Factors influencing the estimated longevity remaining calculation include pacing rate, amplitude, pulse-width and lead impedance.Any even slight changes in these factors will impact the battery consumption calculation and therefore the remaining longevity estimate.Please refer to the description for more information regarding the specific circumstances of this event.Please note this pacemaker's battery was designed to estimate remaining longevity and trigger corresponding device replacement indicators based on the pacing capacitor charge time, which progressively increases as the battery depletes.These pacemakers were not designed with an ability to calculate remaining longevity for every possible current drain, which may cause replacement indicators to trigger earlier than what was previously estimated; however, it does not negatively impact the use, operation, safety, or performance of the device.
|
| |
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that battery status indicators were different than expected based on information provided to boston scientific.Factors influencing the estimated longevity remaining calculation include pacing rate, amplitude, pulse-width and lead impedance.Any even slight changes in these factors will impact the battery consumption calculation and therefore the remaining longevity estimate.Please refer to the description for more information regarding the specific circumstances of this event.Please note this pacemaker's battery was designed to estimate remaining longevity and trigger corresponding device replacement indicators based on the pacing capacitor charge time, which progressively increases as the battery depletes.These pacemakers were not designed with an ability to calculate remaining longevity for every possible current drain, which may cause replacement indicators to trigger earlier than what was previously estimated; however, it does not negatively impact the use, operation, safety, or performance of the device.The device has been received for analysis.Upon completion of the analysis of the complaint device, if there is any further relevant information from that review, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that this device showed half a year remaining from two months ago with magnet rate of ninety.Recently, device battery was at one year remaining with the same magnet rate.Boston scientific technical services (ts) discussed battery performance would vary a lot in this device in which it could be that the low estimate was right after a telemetry session that took longer or lead testing.Ts then advised to save to disk for analysis.To date, the device remains in service.No adverse patient effects reported.Additional information indicated that the device was explanted.No additional adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this device showed half a year remaining from two months ago with magnet rate of ninety.Recently, device battery was at one year remaining with the same magnet rate.Boston scientific technical services (ts) discussed battery performance would vary a lot in this device in which it could be that the low estimate was right after a telemetry session that took longer or lead testing.Ts then advised to save to disk for analysis.To date, the device remains in service.No adverse patient effects reported.Additional information indicated that the device was explanted.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that battery status indicators were different than expected based on information provided to boston scientific.Factors influencing the estimated longevity remaining calculation include pacing rate, amplitude, pulse-width and lead impedance.Any even slight changes in these factors will impact the battery consumption calculation and therefore the remaining longevity estimate.Please refer to the description for more information regarding the specific circumstances of this event.Please note this pacemaker's battery was designed to estimate remaining longevity and trigger corresponding device replacement indicators based on the pacing capacitor charge time, which progressively increases as the battery depletes.These pacemakers were not designed with an ability to calculate remaining longevity for every possible current drain, which may cause replacement indicators to trigger earlier than what was previously estimated; however, it does not negatively impact the use, operation, safety, or performance of the device.The device has been received for analysis.Upon completion of the analysis of the complaint device, if there is any further relevant information from that review, a supplemental report will be filed.The returned pacemaker was analyzed and an engineering-level longevity prediction calculation was completed to assess the rate of battery depletion.Given the programmed parameters and other data stored within the memory of the device, the results of this calculation indicated that the actual rate of battery depletion fell within an acceptable range.Next, a series of diagnostic tests were conducted that verified the performance of pacing, sensing, and recording functions.Having met the engineering longevity prediction, functionally passed all returned product testing, and with no further information to indicate a product performance issue, we have concluded that this device experienced normal battery depletion.
|
| |
|