C.R. BARD, INC. (COVINGTON) -1018233 BARD® URETHRAL CATHETER Back to Search Results
Model Number 120614
Medical Device Problem Code Component Misassembled (4004)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the color of the bard tiemann model coude tip 14fr intermittent catheter was lighter orange when the customer first started using it, but the catheters from the last order was more of a darker shade of orange or red.It was also stated that the darker ones seemed like the rubber was harder than the lighter one and harder to insert which seemed to cause more irritation than the lighter ones.Per additional information received via email on 14jan2022, the problem was the dark brown colored catheters seemed harder than the lighter orange colored ones causing more irritation when inserted.It was also stated that the darker colored ones also had the dimple that did not line up with the curved end, which made it more difficult to insert straight.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the color of the bard tiemann model coude tip 14fr intermittent catheter was lighter orange when the customer first started using it, but the catheters from the last order was more of a darker shade of orange or red.It was also stated that the darker ones seemed like the rubber was harder than the lighter one and harder to insert which seemed to cause more irritation than the lighter ones.Per additional information received via email on 14jan2022, the problem was the dark brown colored catheters seemed harder than the lighter orange colored ones causing more irritation when inserted.It was also stated that the darker colored ones also had the dimple that did not line up with the curved end, which made it more difficult to insert straight.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Corrections: d, g, h.H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the color of the bard tiemann model coude tip 14fr intermittent catheter was lighter orange when the customer first started using it, but the catheters from the last order was more of a darker shade of orange or red.It was also stated that the darker ones seemed like the rubber was harder than the lighter one and harder to insert which seemed to cause more irritation than the lighter ones.Per additional information received via email on 14jan2022, the problem was the dark brown colored catheters seemed harder than the lighter orange colored ones causing more irritation when inserted.It was also stated that the darker colored ones also had the dimple that did not line up with the curved end, which made it more difficult to insert straight.Per additional information via form with sample returned on (b)(6) 2022, customer had urinary tract infection using the intermittent catheter.Customer also stated that the catheters were brown in color and not in orange or red , seem harder.On some catheter the dimples did not match up with the curve on the other end.Customer concerned if they were as sterile as having more urinary tract infection than with the ones that were a different color.The brown ones seem different from the catheter which customer used the last couple of years.The curve not going right way it makes it harder to insert.No medical intervention was reported.
 
Event or Problem Description
It was reported that the color of the bard tiemann model coude tip 14fr intermittent catheter was lighter orange when the customer first started using it, but the catheters from the last order was more of a darker shade of orange or red.It was also stated that the darker ones seemed like the rubber was harder than the lighter one and harder to insert which seemed to cause more irritation than the lighter ones.Per additional information received via email on 14jan2022, the problem was the dark brown colored catheters seemed harder than the lighter orange colored ones causing more irritation when inserted.It was also stated that the darker colored ones also had the dimple that did not line up with the curved end, which made it more difficult to insert straight.Per additional information via form with sample returned on 09feb2022, customer had urinary tract infection using the intermittent catheter.Customer also stated that the catheters were brown in color and not in orange or red , seem harder.On some catheter the dimples did not match up with the curve on the other end.Customer concerned if they were as sterile as having more urinary tract infection than with the ones that were a different color.The brown ones seem different from the catheter which customer used the last couple of years.The curve not going right way it makes it harder to insert.No medical intervention was reported.
 
Additional Manufacturer Narrative
The reported event was confirmed manufacturing related.Samples were confirmed to exhibit the reported failure.A potential root cause for this failure could be "machine error".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.A labeling review is not required as the reported event was found to be manufacturing related.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® URETHRAL CATHETER
Common Device Name
URETHRAL CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13467459
Report Number1018233-2022-00423
Device Sequence Number3886760
Product Code KOD
UDI-Device Identifier00801741029967
UDI-Public(01)00801741029967
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/10/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/10/2026
Device Model Number120614
Device Catalogue Number120614
Device Lot NumberNGFR2541
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/09/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/14/2022
Supplement Date Received by Manufacturer02/09/2022
06/10/2022
Initial Report FDA Received Date02/06/2022
Supplement Report FDA Received Date03/08/2022
06/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-