• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ PANOPTIX TORIC UV ABSORBING TRIFOCAL IOL; LENS, MULTIFOCAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ PANOPTIX TORIC UV ABSORBING TRIFOCAL IOL; LENS, MULTIFOCAL Back to Search Results
Model Number TFAT40
Medical Device Problem Codes Defective Component (2292); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Visual Disturbances (2140)
Date of Event 11/30/2021
Type of Reportable Event Serious Injury
Event or Problem Description
A facility representative reported that following an intraocular lens (iol) implant procedure, the patient experienced vision issues.The iol was exchanged for another lens model 2 months following the initial implant procedure.The clinical reason to explant the lens due to haptic was compromised when facility looked under microscope.Additional information received and stated that, after primary surgery, the patient was brought back to operation room for reposition of lens and noted lens haptic was noted to be compromised at that time.The patient was not hospitalized for the event and there was no patient harm.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The lens was returned for evaluation.Solution is dried on the lens.One haptic is broken in the gusset area and the distal area.Both pieces were returned loose in the lens case.The optic is cut in half, typical of insertion and removal from the eye.A power and resolution inspection could not be conducted due to extensive damage.Broken haptic damage was observed.All lenses are 100% inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met company release criteria.Based on our observation, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ PANOPTIX TORIC UV ABSORBING TRIFOCAL IOL
Common Device Name
LENS, MULTIFOCAL
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key13468861
Report Number1119421-2022-00193
Device Sequence Number4826804
Product Code MFK
UDI-Device Identifier00380652419185
UDI-Public00380652419185
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P040020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/21/2026
Device Model NumberTFAT40
Device Catalogue NumberTFAT40.175
Device Lot Number15169449
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/19/2022
Supplement Date Received by Manufacturer02/21/2022
Initial Report FDA Received Date02/07/2022
Supplement Report FDA Received Date03/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PROVISC OPHTHALMIC VISCOSURGICAL DEVICE
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Patient SexFemale
-
-