| Model Number |
A35HPV06040080 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/02/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Physician was attempting to use a fortrex pta balloon during procedure to treat a little calcified fibrous lesion in the mid av fistula with 80% stenosis.The vessel was moderately tortuous.No embolic protection was used.A medtronic inflation device was used for balloon inflation.There was no damage noted to packaging.There was no issue noted when removing the device from hoop/tray.It was not specified if ifu was followed.The device was prepped per ifu with no issues identified.It was reported that during first balloon inflation a longitudinal burst occurred at 20atm.Not all fragments of the balloon were retrieved.Physician used a non-medtronic stent to pin down remaining balloon material.The device did not pass through a previously deployed stent.There was no resistance encountered when advancing the device.A 10x80 non-medtronic stent was deployed with no issue and procedure was completed.There was no vessel injury reported.No further patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|