| Model Number |
01.32.156DH |
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
01/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Batch review performed on 26 january 2022.Lot 2009438: (b)(4) items manufactured and released on 10-dec-2020.Expiration date: 2025-11-30.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event.Additional implant involved: liner: mpact 01.32.3648hct flat pe hc liner ø36/f (k103721) lot.2013105.Batch review performed on 26 january 2022.Lot 2013105: (b)(4) items manufactured and released on 13-jan-2021.Expiration date: 2026-01-03.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event.Additional implant involved: stem: amistem p 01.18.403 amistem-p std stem size 3 (k173794) lot.2011357.Batch review performed on 26 january 2022.Lot 2011357: (b)(4) items manufactured and released on 27-jan-2021.Expiration date: 2026-01-18.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event.Additional implant involved: ball heads: mectacer 01.29.209 biolox delta ceramic ball head 12/14 ø 36 size m 0 (k112115) lot.2011678.Batch review performed on 26 january 2022.Lot 2011678: (b)(4) items manufactured and released on 22-feb-2021.Expiration date: 2026-02-10.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event.
|
| |
|
Event or Problem Description
|
|
At about 6 months after the primary, the patient came in due to signs of an infection and the pathogen is unknown.The surgeon performed a washout, removed all components, and implanted an antibiotic spacer.The surgery was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Batch review performed on 26 january 2022.Lot 2009438: (b)(4) items manufactured and released on 10-dec-2020.Expiration date: 2025-11-30.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event.Additional implant involved: liner: impact 01.32.3648hct flat pe hc liner ø36/f (k103721) lot.2013105.Batch review performed on 26 january 2022.Lot 2013105: (b)(4) items manufactured and released on 13-jan-2021.Expiration date: 2026-01-03.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event.Additional implant involved: stem: amistem p 01.18.403 amistem-p std stem size 3 (k173794) lot.2011357.Batch review performed on 26 january 2022.Lot 2011357: (b)(4) items manufactured and released on 27-jan-2021.Expiration date: 2026-01-18.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event.Additional implant involved: ball heads: mectacer 01.29.209 biolox delta ceramic ball head 12/14 ø 36 size m 0 (k112115) lot.2011678.Batch review performed on 26 january 2022.Lot 2011678: (b)(4) items manufactured and released on 22-feb-2021.Expiration date: 2026-02-10.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event.
|
| |
|
Event or Problem Description
|
|
At about 6 months after the primary, the patient came in due to signs of an infection and the pathogen is unknown.The surgeon performed a washout, removed all components, and implanted an antibiotic spacer.The surgery was completed successfully.
|
| |
|