Manufacturer ref# (b)(4).Occupation: unknown, but referred to as a health care professional investigation is still in progress.
|
| |
|
Event or Problem Description
|
|
According to initial reporter: a (b)(6) male patient underwent a tvar type b dissection procedure in which the zenith alpha thoracic endovascular graft proximal tapered components, g38178, was used.The trigger wire would not release from the cannula.Physician took the back handle off and pulled the trigger wires off externally.The graft deployed successfully.It was noted that it took a lot of force to pull back on the handle.Patient outcome: the patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer ref# (b)(4).Summary of investigational findings: it was reported by phone a 56 year old male patient underwent a tevar type b dissection procedure in which a zta-pt-36-32-161-w (complaint device) was used.The patient anatomy was reported to be tortuous.The trigger wire would not release from the cannula, so the district manager and physician took the back handle off and pulled the trigger wires off externally.The graft deployed successfully.It was noted that it took a lot of force to pull back on the handle.It was reported that there was no harm to patient.No product was returned and no imaging was provided.Review of the device history record gave no indication of the device being produced out of specification.Per instructions for use: the safety and effectiveness of the zenith alpha thoracic endovascular graft and ancillary components have not been evaluated in the following patient populations: dissections.Futhermore : do not continue advancing the wire guide or any portion of the introduction system if resistance is felt.Stop and assess the cause of resistance; vessel, catheter, or graft damage may occur.Exercise particular care in areas of stenosis, intravascular thrombosis, or calcified or tortuous vessels.Based on the provided information it has not been possible to establish the cause of the reported event.However the tortuos vessels and use outside ifu could have contributed to the reported event.After investigation the event for this pr# is no longer reportable.Cook medical will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
|
| |
|