| Model Number |
STRIP, TF WGN 50CTMG/DL 956970 |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Consumer reported complaint for low and erratic blood glucose test results.The customer is concerned with test results from back to back blood tests of 20 and 92 mg/dl.The customer¿s expected am fasting blood glucose test result range is 75-95 mg/dl and expected pm non-fasting blood glucose test result range is 95-121 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call, a blood test was performed by the customer non-fasting and produced test result of 96 mg/dl using true focus meter; the customer was satisfied with the result obtained.The product is stored according to specification in the living room.The test strip lot manufacturer¿s expiration date is 10/19/2023 and open vial date is (b)(6) 2022.The meter memory was reviewed for previous test result history: (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
Internal report reference number: (b)(4).Meter and test strips were not returned for evaluation.Note: manufacturer contacted customer in a follow-up call on 31-jan-2022 to ensure the replacement products resolved the initial concern - able to establish contact with customer who stated replacement products resolved initial concern.
|
| |
|
Search Alerts/Recalls
|