• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number MI1000 MED-EL CONCERT PIN
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code Implant Pain (4561)
Date of Event 06/11/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Conclusions: according to the limited information from the field, the recipient reports pain with the use of the device.From the device investigation the reason for the reported issue could not be concluded.No obvious device malfunction could be observed, but a possible moisture issue has been noticed during investigation.However, despite further tests a source for the increased moisture could not be found.Mechanical damages found during investigation are attributable to the removal surgery.Hence, the root cause for the reported issues remains unknown.This is a combined initial/final repot.
 
Event or Problem Description
Two months prior the recipient reported that her right implant started buzzing and causing pain.When asked specifics, she pointed to the middle of her coil as the source of the pain.The recipient stopped wearing her right audio processor at that time.External equipment was replaced with no improvement.The recipient was reimplanted and is doing well with her new device.She did not note any pain at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key13470728
Report Number9710014-2022-00088
Device Sequence Number11840966
Product Code MCM
UDI-Device Identifier09008737083625
UDI-Public(01)09008737083625
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/01/2015
Device Model NumberMI1000 MED-EL CONCERT PIN
Device Catalogue Number08841
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2020
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/04/2022
Initial Report FDA Received Date02/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/01/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age12 YR
Patient SexFemale
-
-