• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ADVANIX BILIARY; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ADVANIX BILIARY; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00532190
Medical Device Problem Codes Positioning Failure (1158); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/23/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).(b)(4).The returned advanix biliary stent was analyzed, and a visual evaluation noted that the stent was severely kinked and damaged.No other issues with the device were noted.The reported event was confirmed.Based on the provided and collected information, this device met all manufacturing requirements and no abnormalities were reported during the manufacturing process.After analysis, it was determined that the stent was kinked and damaged, the coils of the stent failed to curl.It's most likely that the kinks and damages observed on the stent may be related to user manipulation while preparing the stent and/or anatomical factors could have contributed on this issue during the procedure, also it's possible that the anatomical factors plus the technique performed during procedure caused the issue of the coils failing to curl or maintain the correct shape, in summary those conditions on the stent avoided the correct deployment of it.Therefore, of all available information indicated that the root cause of the problems found is adverse event related to procedure since the adverse event occurred during the procedure and the device had no influence on event.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Event or Problem Description
It was reported boston scientific corporation that an advanix biliary stent was used during an obstructive jaundice procedure in the lower common bile duct performed on (b)(6) 2021.During the procedure, when the advanix biliary stent was placed in the bile duct, it became kinked after withdrawing the push catheter.It was noted that they had difficulty separating the advanix biliary stent from the push catheter upon reaching the end of the stent.The procedure was successfully completed with another advanix biliary stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.Investigation results revealed that the stent was damaged, therefore this event has been deemed an mdr reportable event.Please see for full investigation details.
 
Additional Manufacturer Narrative
(b)(6).(b)(4).The returned advanix biliary stent was analyzed, and a visual evaluation noted that the stent was severely kinked and damaged.No other issues with the device were noted.The reported event was confirmed.Based on the provided and collected information, this device met all manufacturing requirements and no abnormalities were reported during the manufacturing process.After analysis, it was determined that the stent was kinked and damaged, the coils of the stent failed to curl.It's most likely that the kinks and damages observed on the stent may be related to user manipulation while preparing the stent and/or anatomical factors could have contributed on this issue during the procedure, also it's possible that the anatomical factors plus the technique performed during procedure caused the issue of the coils failing to curl or maintain the correct shape, in summary those conditions on the stent avoided the correct deployment of it.Therefore, of all available information indicated that the root cause of the problems found is adverse event related to procedure since the adverse event occurred during the procedure and the device had no influence on event.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Event or Problem Description
It was reported boston scientific corporation that an advanix biliary stent was used during an obstructive jaundice procedure in the lower common bile duct performed on (b)(6) 2021.During the procedure, when the advanix biliary stent was placed in the bile duct, it became kinked after withdrawing the push catheter.It was noted that they had difficulty separating the advanix biliary stent from the push catheter upon reaching the end of the stent.The procedure was successfully completed with another advanix biliary stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.Investigation results revealed that the stent was damaged, therefore this event has been deemed an mdr reportable event.Please see for full investigation details.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANIX BILIARY
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13470844
Report Number3005099803-2022-00301
Device Sequence Number17222119
Product Code FGE
UDI-Device Identifier08714729786986
UDI-Public08714729786986
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K101786
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/05/2022
Device Model NumberM00532190
Device Catalogue Number3219
Device Lot Number0025814250
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/06/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/10/2022
Initial Report FDA Received Date02/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/05/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age62 YR
Patient SexUnknown
Patient Weight70 KG
-
-