| Model Number |
M00532190 |
| Medical Device Problem Codes |
Positioning Failure (1158); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).(b)(4).The returned advanix biliary stent was analyzed, and a visual evaluation noted that the stent was severely kinked and damaged.No other issues with the device were noted.The reported event was confirmed.Based on the provided and collected information, this device met all manufacturing requirements and no abnormalities were reported during the manufacturing process.After analysis, it was determined that the stent was kinked and damaged, the coils of the stent failed to curl.It's most likely that the kinks and damages observed on the stent may be related to user manipulation while preparing the stent and/or anatomical factors could have contributed on this issue during the procedure, also it's possible that the anatomical factors plus the technique performed during procedure caused the issue of the coils failing to curl or maintain the correct shape, in summary those conditions on the stent avoided the correct deployment of it.Therefore, of all available information indicated that the root cause of the problems found is adverse event related to procedure since the adverse event occurred during the procedure and the device had no influence on event.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
|
| |
|
Event or Problem Description
|
|
It was reported boston scientific corporation that an advanix biliary stent was used during an obstructive jaundice procedure in the lower common bile duct performed on (b)(6) 2021.During the procedure, when the advanix biliary stent was placed in the bile duct, it became kinked after withdrawing the push catheter.It was noted that they had difficulty separating the advanix biliary stent from the push catheter upon reaching the end of the stent.The procedure was successfully completed with another advanix biliary stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.Investigation results revealed that the stent was damaged, therefore this event has been deemed an mdr reportable event.Please see for full investigation details.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(6).(b)(4).The returned advanix biliary stent was analyzed, and a visual evaluation noted that the stent was severely kinked and damaged.No other issues with the device were noted.The reported event was confirmed.Based on the provided and collected information, this device met all manufacturing requirements and no abnormalities were reported during the manufacturing process.After analysis, it was determined that the stent was kinked and damaged, the coils of the stent failed to curl.It's most likely that the kinks and damages observed on the stent may be related to user manipulation while preparing the stent and/or anatomical factors could have contributed on this issue during the procedure, also it's possible that the anatomical factors plus the technique performed during procedure caused the issue of the coils failing to curl or maintain the correct shape, in summary those conditions on the stent avoided the correct deployment of it.Therefore, of all available information indicated that the root cause of the problems found is adverse event related to procedure since the adverse event occurred during the procedure and the device had no influence on event.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
|
| |
|
Event or Problem Description
|
|
It was reported boston scientific corporation that an advanix biliary stent was used during an obstructive jaundice procedure in the lower common bile duct performed on (b)(6) 2021.During the procedure, when the advanix biliary stent was placed in the bile duct, it became kinked after withdrawing the push catheter.It was noted that they had difficulty separating the advanix biliary stent from the push catheter upon reaching the end of the stent.The procedure was successfully completed with another advanix biliary stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.Investigation results revealed that the stent was damaged, therefore this event has been deemed an mdr reportable event.Please see for full investigation details.
|
| |
|