BIOTRONIK AG, BUELACH, SWITZERLAND PASSEO-35 8/40/80; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 357291
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/11/2022
Type of Reportable Event Malfunction
Event or Problem Description
A passeo-35 balloon catheter was selected for treatment of an aneurysm in the ostial subclavian artery.The device was advance over the wire and inflated at subclavian artery to post dilate a covered stent.11 bar pressure was applied.The balloon could not inflate further and leaking of contrast was observed in the fluoroscopy.Blood also observed in the syringe of the inflation device.
 
Additional Manufacturer Narrative
The returned instrument was subjected to a technical analysis and the corresponding product release documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation of the returned device revealed a tear with a length of 1 mm, which penetrates the balloon wall 22 mm proximal to the distal x-ray marker.It seems likely that the damage of the balloon surface was caused by a hard, sharp-edged object such as e.G.Anatomical structure.Review of the product release documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each instrument is tested for air tightness by means of a pressure test and a helium leak test.We can therefore confirm that the instrument was delivered in a leak-proof condition.Based on the conducted investigations, no manufacturing or material related root cause could be determined.The root cause is most likely related to the patients anatomy.It should be noted that the ifu contraindicates the treatment of lesions that lies within or adjacent to an aneurysm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PASSEO-35 8/40/80
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key13471102
Report Number1028232-2022-00627
Device Sequence Number4128387
Product Code DQY
UDI-Device Identifier07640119558202
UDI-Public07640119558202
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K082933
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2024
Device Model Number357291
Device Catalogue NumberSEE MODEL NO.
Device Lot Number03210827
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/02/2022
Supplement Date Received by Manufacturer03/28/2022
Initial Report FDA Received Date02/07/2022
Supplement Report FDA Received Date04/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-