| Model Number |
PLC141000 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Codes |
Pain (1994); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
01/13/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).This event also includes device 21562000/ 00733132618767.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2021, the patient was treated for an abdominal aortic aneurysm.The physician implanted a gore® excluder® iliac branch endoprosthesis and two gore® excluder® aaa endoprostheses.The patient tolerated the procedure.On (b)(6) 2022, the patient was seen for a followup exam.The patient reported new onset left leg pain and buttock claudication.It was determined that the issues were due to the left iliac limb being compressed due to a tight distal aorta.The physician implanted a covered stent on (b)(6) 2022.The patient tolerated the procedure.
|
| |
|