| Model Number |
CA2110X12B |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer received information alleging a ventilator inoperative condition occurred.There was no harm or injury reported.The device has not yet been returned to the manufacturer for evaluation.At this time, we are unable to confirm the alleged malfunction.A follow-up report will be submitted when the manufacturer's investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
It was initially reported, the manufacturer received information alleging a ventilator inoperative condition occurred.There was no harm or injury reported.During the evaluation of the device at the manufacturer's service center, a "ventilator inoperative" code was found in the ventilator's downloaded error log.The device's system board was replaced to address the issue.
|
| |
|