Event or Problem Description
The manufacturer received information alleging a ventilator inoperative condition occurred.There was no harm or injury reported.The device has not yet been returned to the manufacturer for evaluation.At this time, we are unable to confirm the alleged malfunction.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Additional Manufacturer Narrative
It was initially reported, the manufacturer received information alleging a ventilator inoperative condition occurred.There was no harm or injury reported.During the evaluation of the device at the manufacturer's service center, a "ventilator inoperative" code was found in the ventilator's downloaded error log.The device's system board was replaced to address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY EVO
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15206
MDR Report Key13471610
Report Number2518422-2022-03299
Device Sequence Number4870856
Product Code CBK
UDI-Device Identifier00606959052307
UDI-Public00606959052307
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCA2110X12B
Device Catalogue NumberCA2110X12B
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/11/2022
Supplement Date Received by Manufacturer04/06/2022
Initial Report FDA Received Date02/07/2022
Supplement Report FDA Received Date04/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/29/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1