A supplemental report is being submitted for device evaluation and investigation completion.Product event summary: the controller ac adapter was returned for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.Functional testing revealed that the controller ac adapter was unable to provide power.Supplemental testing revealed damage to the wires within the connector's strain relief; further analysis revealed that open strands from the damaged wires came in contact with each other, resulting in a short circuit condition.After the strands from the damaged wires were separated, the device regained functionality.As a result, the reported event was confirmed.The most likely root cause of the reported event can be attributed to wear on the strain relief and/or the handling of the device, which likely contributed to the breaking of the cable wires.Capa pr00405633 is investigating damage to power sources.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|