The following fields were updated due to corrected information: h.6.Imdrf annex a grid: a0504 - leak / splash the following fields were updated due to additional information: d.9.Device available for eval?: yes.D.9.Returned to manufacturer on: 03-mar-2022.H.6.Investigation: two samples and one photo were received by our quality team for evaluation.From the photo, a 22g venflon pro safety product with blood leakage from the end cap was observed.A representative sample from batch 1140524 was received.The sample was subjected to visual inspection.The sample passed the acceptance criteria, and no abnormalities were observed.The end cap belongs to cavity #21.No hole was observed on the end cap luer cone tip.One used (contaminated) end cap with cavity #49 was also received.The used end cap was subjected to visual inspection.A sink mark was observed on the end cap and a through hole was observed on the luer cone tip.A review of the internal manufacturing device records and raw material history files for the reported lot number was performed and no recorded quality problems or rejections to this incident were found.Based on the end cap batch usage record, for the reported batch 1140524, cavity #49 was not used in the production of this assembled batch.Cavity #49 could be a sample from the same customer of another assembly batch.The probable root cause for the reported leakage could be due to the sink mark hole at the luer cone tip of the end cap.The air vent may be choked during molding and cause hot air to be trapped.This affects the material flow to the final fill area at the luer cone tip.A project, capa#4710247, has been initiated to address this issue.H3 other text : see h.10.
|