Model Number ACH4041-01
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/11/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that during a fasciectomy procedure, the tip of the topaz wand broke off inside the patient.No broken piece inside the patient was located.The procedure was successfully completed without delay using a back-up device.No patient complications were reported.
 
Event or Problem Description
It was reported that during a fasciectomy procedure, the tip of the topaz wand broke off inside the patient.All the pieces were removed.Upon removal from the patient the tip was noted to be broken and could not be located.A flat plate was completed and read by the radiologist and confirmed to not be retained.The procedure was successfully completed without delay using a back-up device.No further complications were reported.
 
Additional Manufacturer Narrative
H10: h3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states based on the information provided, ¿upon removal from the patient the tip was noted to be broken and could not be located.A flat plate was completed and read by the radiologist and confirmed to not be retained.Staff looked throughout the drapes, the floor, and trash but did not locate the tip to the topaz device.¿ these devices are manufactured and intended as externally communicating devices and are not approved for long-term internal tissue exposure and long-term implantation data is not available.The patient impact beyond possible micro-motion and/or migration, and local irritation/discomfort cannot be determined.According to the report, the procedure was successfully completed without delay using a back-up device.No patient complications were reported.Should any additional relevant patient information be provided, this case would be re-assessed.There was no relationship found between the device and the reported event.The complaint was not confirmed.Factors that could have contributed to the reported event include: (1) hitting the device tip against a hard surface.(2) using the device as a lever to enlarge surgical site.(3) mechanical displacement of tissue through applied force.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOPAZ EZ IFS
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13471936
Report Number3006524618-2022-00049
Device Sequence Number20051252
Product Code GEI
UDI-Device Identifier00817470003574
UDI-Public00817470003574
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K140521
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/01/2023
Device Model NumberACH4041-01
Device Catalogue NumberACH4041-01
Device Lot Number2061034
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/18/2022
Supplement Date Received by Manufacturer11/24/2022
Initial Report FDA Received Date02/07/2022
Supplement Report FDA Received Date11/29/2022
Date Device Manufactured10/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
-
-