| Model Number |
H7493941816220 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/09/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: synergy xd mr us 2.25 x 16mm stent delivery system was returned for analysis.Examination of the crimped stent via scope found no issues with the stent.There were no signs of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal marker bands.The crimped stent outer diameter was measured and the result was within max crimped stent profile measurement.The balloon cones were reviewed via scope, and no issues were noted.The balloon wings were tightly wrapped and evenly folded.The balloon had not been subjected to positive pressure.A visual and tactile examination of the hypotube found multiple links along the hypotube.A visual and microscopic examination identified a partial fracture, 34.6cm proximal from the tip, at the lasercut region.Examination of the tip via scope found no damage to the tip.The shaft polymer extrusion including the distal and mid-shaft sections were examined via scope and a tactile examination was performed.No issues were noted.An attempt was made to inflate the balloon using an encore inflation unit.The balloon failed to inflate due to liquid leaking out through the partial fracture at the lasercut region.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 26jan2022.It was reported that the balloon inflation failure occurred.The 80% stenosed target lesion was located in the mildly tortuous and mildly calcified diagonal branch.A 2.25 x 16mm synergy xd drug-eluting stent was advanced for treatment.However, during inflation, the balloon would not inflate.The stent was removed and the it was noticed that there was blood in the inflator.The procedure was completed with a 2.25x12mm synergy xd drug-eluting stent.No patient complications were reported.However, returned device analysis revealed a partial fracture of the shaft.
|
| |
|
Search Alerts/Recalls
|