Model Number SCRWDRV SFT,SHT NON-CANN,2.5MM HEX
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 01/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
On 01/20/2022 it was reported by a sales representative via sems that an ar-1995shn was used in implant removal surgery after acl reconstruction on (b)(6) 2021.When the screw was removed, the product was broken, and the broken piece fell off inside the screw.The broken piece and the screw could not be removed and the surgery was completed with the screw left in the body.
 
Additional Manufacturer Narrative
Complaint confirmed.One unpackaged ar-1995shn was received for investigation.Visual inspection identified that the distal tip of the drive geometry had sheared off, with the breakage site noticeably twisted.Assessment with digital micrometer id: 224 identified that the driver shaft met design specifications.The probable cause is attributed to over-torquing/over-engaging the driver within the screw head during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCRWDRV SFT,SHT NON-CANN,2.5MM HEX
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key13472179
Report Number1220246-2022-04431
Device Sequence Number1763923
Product Code LXH
UDI-Device Identifier00888867025684
UDI-Public00888867025684
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 04/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSCRWDRV SFT,SHT NON-CANN,2.5MM HEX
Device Catalogue NumberAR-1995SHN
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/19/2022
Supplement Date Received by Manufacturer01/19/2022
Initial Report FDA Received Date02/07/2022
Supplement Report FDA Received Date04/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-