MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number 3542757MC
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Codes Capsular Contracture (1761); Granuloma (1876); Seroma (2069)
Date of Event 06/03/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The complaint device has been discarded.As a result, no product failure analysis can be conducted, and device malfunction cannot be confirmed.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: right rupture, granuloma; late onset seroma, and capsular contracture; baker grade iii.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a female patient underwent revision breast augmentation with 275cc mentor memorygel breast implant on the right side, and with 300cc mentor memorygel breast implant on the left (date of implant: (b)(6) 2014) and experienced breast implant rupture on the right side complicated by granuloma formation in the right axilla region confirmed by biopsy performed during surgery.It was also reported that the patient developed capsular contracture; baker grade ii on the left, and baker grade iii on the right.It was also reported that the patient developed late onset seroma, the fluid was sent for pathology, at this time, the results of the pathology report have not been provided.As a result, the patient underwent bilateral explantation and replacement with bilateral new saline implants on (b)(6) 2022.This medwatch report is for the patient¿s right-sided device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
gabriel alfageme
3041 skyway circle north
irving, TX 75038
9497898687
MDR Report Key13472409
Report Number1645337-2022-01267
Device Sequence Number17349998
Product Code FTR
UDI-Device Identifier00081317004387
UDI-Public00081317004387
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2016
Device Model Number3542757MC
Device Catalogue Number3542757MC
Device Lot Number6534572
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/17/2022
Initial Report FDA Received Date02/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/19/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age50 YR
Patient SexFemale
-
-