IDEAL IMPLANT INCORPORATED IDEAL IMPLANT STRUCTURED BREAST IMPLANT; SALINE-FILLED BREAST IMPLANT Back to Search Results
Model Number 21001
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 12/01/2021
Type of Reportable Event Serious Injury
Event or Problem Description
Alleged deflation.
 
Additional Manufacturer Narrative
No manufacturing defect was found on macroscopic or microscopic examination of the explanted device.A crease fold failure was observed on the outer shell resulting in a leak.Updates made to sections b5, d6b, d9, g3, g6, h2, h3, h6, h10.
 
Event or Problem Description
Deflation resulting in explantation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IDEAL IMPLANT STRUCTURED BREAST IMPLANT
Common Device Name
SALINE-FILLED BREAST IMPLANT
Manufacturer (Section D)
IDEAL IMPLANT INCORPORATED
14881 quorum drive
suite 925
dallas TX 75254
Manufacturer (Section G)
SPECIALTY SILICONE FABRICATORS INC.
2761 walnut avenue
tustin CA 92780
Manufacturer Contact
robert hamas
14881 quorum drive
suite 925
dallas, TX 75254
MDR Report Key13472682
Report Number3011491947-2022-00031
Device Sequence Number17316103
Product Code FWM
UDI-Device Identifier10851795006008
UDI-Public(01)10851795006008(241)21001(17)200715
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P120011
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/15/2020
Device Model Number21001
Device Catalogue Number210
Device Lot Number62121
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/03/2022
Supplement Date Received by Manufacturer04/04/2022
Initial Report FDA Received Date02/07/2022
Supplement Report FDA Received Date04/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/26/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age50 YR
Patient SexFemale
-
-