• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROPORT ORTHOPEDICS INC. PROFEMUR® MODULAR FEMORAL NECK; HIP COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROPORT ORTHOPEDICS INC. PROFEMUR® MODULAR FEMORAL NECK; HIP COMPONENT Back to Search Results
Model Number PHA01254
Medical Device Problem Codes Break (1069); Difficult to Remove (1528)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This event will be updated once the investigation is complete.Trends will be evaluated.
 
Event or Problem Description
Allegedly, ti modular neck broke when pt picked up a potted plant.Attempted to remove with mpo remnant removal tray, 2 drill bits and a tap broke in the process.Surgeon also attempted to bur out neck with metal bur, this also did not work.Surgeon also attempted to remove with osteotomes, this did not work.After 4 hours the surgeon was finally able to remove the stem by performing an eto.Implants not revised: partid: conserve® plus cup 56mm id 62mm od beaded,item#38025662, lot#087457365, qty:1.
 
Additional Manufacturer Narrative
This event will be updated once the investigation is complete.Trends will be evaluated.
 
Event or Problem Description
Allegedly, ti modular neck broke when pt picked up a potted plant.Attempted to remove with mpo remnant removal tray, 2 drill bits and a tap broke in the process.Surgeon also attempted to bur out neck with metal bur, this also did not work.Surgeon also attempted to remove with osteotomes, this did not work.After 4 hours the surgeon was finally able to remove the stem by performing an eto.Implants not revised: partid: conserve® plus cup 56mm id 62mm od beaded,item#38025662, lot#087457365, qty:1.
 
Additional Manufacturer Narrative
A minor change in section h.3, in the initial report the selection "no" was checked and changed to no returned to manufacturer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROFEMUR® MODULAR FEMORAL NECK
Common Device Name
HIP COMPONENT
Manufacturer (Section D)
MICROPORT ORTHOPEDICS INC.
5677 airline rd.
arlington TN 38002
Manufacturer (Section G)
MICROPORT ORTHOPEDICS INC.
5677 airline rd.
arlington TN 38002
Manufacturer Contact
5677 airline road
arlington, TN 38002
9018674771
MDR Report Key13473311
Report Number3010536692-2022-00043
Device Sequence Number2084995
Product Code LWJ
UDI-Device IdentifierM684PHA012541
UDI-PublicM684PHA012541
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 03/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Model NumberPHA01254
Device Catalogue NumberPHA01254
Device Lot Number086371028
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date01/18/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/18/2022
Supplement Date Received by Manufacturer01/18/2022
Initial Report FDA Received Date02/07/2022
Supplement Report FDA Received Date03/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-2941-2020
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
Patient SexMale
Patient Weight159 KG
-
-