Model Number 9735668
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A medtronic representative went to the site to test the equipment.Testing revealed that hardware was replaced.The system then passed the system checkout and was found to be fully functional.H3: analysis found on emitter- analysis determined the returned units' port six has an amber led light which constantly stays lit, with or without an instrument inserted.Lat does not recognize that an instrument has been inserted.Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: 9735825rpend, serial/lot #: 603004140, ubd: , udi#: 00643169993778.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
(b)(6) 2021 salesforce (b)(4) (rep): medtronic received information regarding a navigation system being used outside of a procedure.It was reported that the site had an amber light on their interface box port 6.  when opening the tracking details, port 6 was listed as faulted.This was noted during setup.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION¿ S8 SYSTEM
Common Device Name
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key13473401
Report Number1723170-2022-00159
Device Sequence Number3963317
Product Code HAW
UDI-Device Identifier00763000359928
UDI-Public00763000359928
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K162309
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9735668
Device Catalogue Number9735668
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/18/2022
Initial Report FDA Received Date02/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/27/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-