(b)(4).Information regarding patient identifier, date of birth, age, gender, weight, race, and ethnicity were not provided.Procode is krd/hcg.The healthcare professional reported that during a coil embolization of an internal iliac artery (iia) aneurysm, the first and second frame s coils were successfully implanted at the target lesion.The complaint coil, an 8mm x 30cm micrusframe 18 (mfr180830 / 30649225) was positioned in the aneurysm as the third framing coil, but the coil failed to detach.The physician pressed the detach button on the control cable and the buzzer sounded.Using fluoroscopic guidance, the physician pulled the device positioning unit (dpu) wire to verify coil detachment.It was recognized that the coil-tail was in a state where it was not able to be pulled together with the wire, and it was not able to be detached.The coil was pulled back to some extent, and the coil then detached.The coil remained inside the concomitant prowler select plus microcatheter (unknown product code and lot number).With the dpu wire, the coil was pushed out of the microcatheter and somehow implanted at the target aneurysm.The procedure progressed, and the stent graft was implanted without incident.There was no report of patient injury.The complaint device is not available for evaluation.Based on complaint information, the device remains implanted and is thus not available for evaluation.A review of manufacturing documentation associated with this lot (30649225) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no nonconformances related to device manufacture or inspection.All product rejected during manufacturing was identified as scrap and properly accounted for.Product analysis cannot be conducted as the product was not returned for analysis.No determination of causes and possible contributing factors could be made.As such, the investigation will be closed.Failure of the coil to detach and unintended detachment during retrieval requiring additional intervention are known potential complications associated with the use of micrusframe 18 in coil embolization procedures.Factors that can contribute to detachment issues include failure to confirm secure connection of the cable to the dpu and detachment control box (dcb).The instructions for use (ifu) advises the user to ensure that the connecting cable is fully seated with the connector end of the dpu wire and output connector of the dcb.The user should verify that the microcoil delivery system is fully connected and that no faults are indicated on the dcb.If a fault exists, all connections should be reseated between the dpu, the dcb, and the connecting cable.If a fault still persists, the connecting cable should be replaced.If this does not correct the error, the dcb should be replaced.If the microcoil delivery system still continues to have a fault, the microcoil should be retrieved and replaced with a new microcoil system.If after depressing the detach button on the dcb, the dcb should be replaced if the light and audible tone do not activate.Also, if the system ready light is not illuminated and there is no detachment after two detachment attempts, the dcb should be replaced.If detachment still does not occur, the microcoil system should be carefully withdrawn and replaced.Following failed coil detachment, it appears that the coil unintendedly detached in the microcatheter during retrieval.Assignment of root cause for the event remains speculative and inconclusive without the return of the associated devices.However, there are clinical and procedural factors that may have contributed rather than the design or manufacture of the device.Since the event required additional intervention (i.E., use of the dpu wire) to preclude patient complications such as non-target site embolization, ischemia, or infarct, the event meets mdr reporting criteria.Based on the device history record review, there is no indication that the event is related to the device manufacturing process.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|
|
The healthcare professional reported that during a coil embolization of an internal iliac artery (iia) aneurysm, the first and second frame s coils were successfully implanted at the target lesion.The complaint coil, an 8mm x 30cm micrusframe 18 (mfr180830 / 30649225) was positioned in the aneurysm as the third framing coil, but the coil failed to detach.The physician pressed the detach button on the control cable and the buzzer sounded.Using fluoroscopic guidance, the physician pulled the device positioning unit (dpu) wire to verify coil detachment.It was recognized that the coil-tail was in a state where it was not able to be pulled together with the wire, and it was not able to be detached.The coil was pulled back to some extent, and the coil then detached.The coil remained inside the concomitant prowler select plus microcatheter (unknown product code and lot number).With the dpu wire, the coil was pushed out of the microcatheter and somehow implanted at the target aneurysm.The procedure progressed, and the stent graft was implanted without incident.There was no report of patient injury.The complaint device is not available for evaluation.
|
|
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to include the additional event information received on 07-feb-2022.E.1: the initial reporter phone: (b)(6).The initial reporter email address is not available / reported.[additional information]: on 07-feb-2022, additional information was received.The information confirmed that the coil embolization was that of the iia; it was an endovascular treatment of abdominal aortic aneurysm (evar) procedure, other details were not reported.A pre-deployment electrical check was performed.A fault light was not seen during the case.Upon pressing the power button, all the lights illuminated.The green system ready light illuminated and during the detachment cycle, the detachment light illuminated and the audible signal beep was heard.All connections fit properly.The reported issue did not result in any clinically significant delay in the procedure.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|