| Model Number |
ORT300 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/08/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During routine preventive maintenance on november 8, 2019, an imris service engineer observed a small seepage leak in the hydraulic line to the back tilt cylinder in the ort300 table.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The pinhole leak occurred in the braided ptfe hydraulic hose at the edge of a hydraulic fitting.The probable cause was due to a worn-out hose against the edge of the fitting.The hydraulic hose was replaced and tested to verify proper function.This mdr is being submitted outside of the required timeframe as part of remedial action initiated by the manufacturer, in response to internally identified issues regarding failed electronic submissions through webtrader.
|
| |
|