• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMRIS - DEERFIELD IMAGING, INC IMRIS ORT300 OPERATING ROOM TABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMRIS - DEERFIELD IMAGING, INC IMRIS ORT300 OPERATING ROOM TABLE Back to Search Results
Model Number ORT300
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/08/2019
Type of Reportable Event Malfunction
Event or Problem Description
During routine preventive maintenance on november 8, 2019, an imris service engineer observed a small seepage leak in the hydraulic line to the back tilt cylinder in the ort300 table.There was no patient involvement.
 
Additional Manufacturer Narrative
The pinhole leak occurred in the braided ptfe hydraulic hose at the edge of a hydraulic fitting.The probable cause was due to a worn-out hose against the edge of the fitting.The hydraulic hose was replaced and tested to verify proper function.This mdr is being submitted outside of the required timeframe as part of remedial action initiated by the manufacturer, in response to internally identified issues regarding failed electronic submissions through webtrader.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMRIS ORT300 OPERATING ROOM TABLE
Common Device Name
OPERATING ROOM TABLE
Manufacturer (Section D)
IMRIS - DEERFIELD IMAGING, INC
5101 shady oak road
minnetonka MN 55343 4100
Manufacturer (Section G)
IMRIS - DEERFIELD IMAGING, INC
5101 shady oak road
minnetonka MN 55343 4100
Manufacturer Contact
kwaku amoah
5101 shady oak road
minnetonka, MN 55343-4100
7632036344
MDR Report Key13476178
Report Number3010326005-2019-00037
Device Sequence Number20120575
Product Code KXJ
UDI-Device Identifier00857534006356
UDI-Public00857534006356
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Biomedical Engineer
Remedial Action Inspection,Repair
Type of Report Initial
Report Date (Section B) 11/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberORT300
Device Catalogue Number114093-000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2019
Initial Report FDA Received Date02/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/09/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-