IMRIS - DEERFIELD IMAGING, INC IMRIS ORT300 OPERATING ROOM TABLE Back to Search Results
Model Number ORT300
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/08/2019
Event Type  malfunction  
Event Description
During routine preventive maintenance on november 8, 2019, an imris service engineer observed a small seepage leak in the hydraulic line to the back tilt cylinder in the ort300 table.There was no patient involvement.
 
Manufacturer Narrative
The pinhole leak occurred in the braided ptfe hydraulic hose at the edge of a hydraulic fitting.The probable cause was due to a worn-out hose against the edge of the fitting.The hydraulic hose was replaced and tested to verify proper function.This mdr is being submitted outside of the required timeframe as part of remedial action initiated by the manufacturer, in response to internally identified issues regarding failed electronic submissions through webtrader.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMRIS ORT300 OPERATING ROOM TABLE
Type of Device
OPERATING ROOM TABLE
Manufacturer (Section D)
IMRIS - DEERFIELD IMAGING, INC
5101 shady oak road
minnetonka MN 55343 4100
Manufacturer (Section G)
IMRIS - DEERFIELD IMAGING, INC
5101 shady oak road
minnetonka MN 55343 4100
Manufacturer Contact
kwaku amoah
5101 shady oak road
minnetonka, MN 55343-4100
7632036344
MDR Report Key13476178
MDR Text Key286449830
Report Number3010326005-2019-00037
Device Sequence Number1
Product Code KXJ
UDI-Device Identifier00857534006356
UDI-Public00857534006356
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Biomedical Engineer
Remedial Action Inspection
Type of Report Initial
Report Date 11/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberORT300
Device Catalogue Number114093-000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/08/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/09/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-