• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMRIS - DEERFIELD IMAGING, INC IMRIS ORT300 OPERATING ROOM TABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMRIS - DEERFIELD IMAGING, INC IMRIS ORT300 OPERATING ROOM TABLE Back to Search Results
Model Number ORT300
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/11/2019
Event Type  malfunction  
Manufacturer Narrative
An imris customer service engineer (cse) was on-site to resolve the issue.The entire table was removed from the customer site and returned for further evaluation.The probable root cause of the table drift was traced to contaminants in the hydraulic fluid.This mdr is being submitted outside of the required timeframe as part of remedial action initiated by the manufacturer, in response to internally identified issues regarding failed electronic submissions through webtrader.
 
Event Description
During an open brain tumor procedure on (b)(6) 2019, the ort300 operating room table drifted in the left roll position at approximately 8-10 degrees.There was a 20 minute delay in the case while the operating room (or) staff attempted to fit the patient into the scanner without compromising the field or patient position.The operating room staff was able to counterbalance the roll movement with two people leaning on the right side of the or table to keep it leveled while the magnet moved over the patient.There was no patient harm or user injury reported due to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMRIS ORT300 OPERATING ROOM TABLE
Type of Device
OPERATING ROOM TABLE
Manufacturer (Section D)
IMRIS - DEERFIELD IMAGING, INC
5101 shady oak road
minnetonka MN 55343 4100
Manufacturer (Section G)
IMRIS - DEERFIELD IMAGING, INC
5101 shady oak road
minnetonka MN 55343 4100
Manufacturer Contact
kwaku amoah
5101 shady oak road
minnetonka, MN 55343-4100
7632036344
MDR Report Key13476371
MDR Text Key286449265
Report Number3010326005-2019-00034
Device Sequence Number1
Product Code KXJ
UDI-Device Identifier00857534006462
UDI-Public00857534006462
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Administrator/Supervisor
Remedial Action Inspection
Type of Report Initial
Report Date 10/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberORT300
Device Catalogue Number114093-600
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2019
Date Manufacturer Received10/11/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/19/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age60 YR
Patient Weight73 KG
-
-