| Model Number |
ORT300 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/11/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An imris customer service engineer (cse) was on-site to resolve the issue.The cse removed and replaced the roll cylinder in the or table and the cylinder was returned for evaluation.During inspection, it was discovered that the cylinder was manufactured from an old design where the o-ring seal was of a lower durometer material resulting in decreased durability.This mdr is being submitted outside of the required timeframe as part of remedial action initiated by the manufacturer, in response to internally identified issues regarding failed electronic submissions through webtrader.
|
| |
|
Event or Problem Description
|
|
On (b)(6), 2019, the customer reported that the ort300 table drifted left.The event occurred during daily qa and there was no patient involvement.
|
| |
|