Model Number 75023360
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/17/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, during a thr surgery, a polarcup trial insert nav 22/51 broke while impacting.The procedure was completed, with a delay less then 30 min, using a s+n back-up device.Patient was not harmed.
 
Additional Manufacturer Narrative
It was reported that, during a thr surgery, a polarcup trial insert nav 22/51 broke while impacting.The procedure was completed with a delay less than 30 min using a s+n back-up device.Patient was not harmed.The complaint device, used in treatment, was not returned for investigation hence the product evaluation could not be performed.A review of the batch record revealed no deviations from the standard manufacturing process.A review of the complaint history revealed no additional complaint for the batch in question.A review of the risk management documentation verifies the failure mode and severity of the reported issue.No prior escalation actions are related to the reported issue.The ifu (no.12.23, ed.01/21) refers to the surgical technique for the correct handling of surgical instruments.The functionality of instruments should be checked and using damaged instruments may lead to early failure to implants.However, in this case it is unclear what may have caused the failure mode.Based on the performed investigations, the reported failure mode could not be confirmed.A relationship between the reported event and the device cannot be confirmed.The reported device met manufacturing specifications upon release for distribution.No probable cause can be determined.Normal wear and tear through repeated use is known to contribute to the reported event.According to document "processing (cleaning, disinfection and sterilization) of instruments from smith & nephew orthopaedics ag" (lit.N°03389-en 1363 v3 11/19), all devices must be inspected and controlled for proper functioning after cleaning/disinfection.This version of the device will be monitored for similar issues.This investigation is considered closed.Should the complaint device become available, the complaint will be reassessed.
 
Additional Manufacturer Narrative
It was reported that, during a thr surgery, a polarcup trial insert nav 22/51 broke while impacting.The procedure was completed, with a delay less than 30 min, using a s+n back-up device.Patient was not harmed.The complaint device, used in treatment, was returned for investigation and the reported failure mode could be confirmed upon visual inspection.A review of the batch record revealed no deviations from the standard manufacturing process.A review of the complaint history revealed no additional complaint for the batch in question.A review of the risk management documentation verifies the failure mode and severity of the reported issue.No prior escalation actions are related to the reported issue.The ifu (no.12.23, ed.03/21) refers to the surgical technique for the correct handling of surgical instruments.The functionality of instruments should be checked and using damaged instruments may lead to early failure to implants.However, in this case it is unclear what may have caused the failure mode.Based on the performed investigations, the reported failure mode could be confirmed.A relationship between the reported event and the device can be confirmed.The reported device met manufacturing specifications upon release for distribution.No probable cause can be determined.Normal wear and tear through repeated use is known to contribute to the reported event.According to document "processing (cleaning, disinfection and sterilization) of instruments from smith & nephew orthopaedics ag" (lit.N°03389-en 1363 v3 11/19), all devices must be inspected and controlled for proper functioning after cleaning/disinfection.This version of the device will be monitored for similar issues.This investigation is considered closed.The complaint device will be discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARCUP TRIAL INSERT NAV 22/51
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13476407
Report Number9613369-2022-00020
Device Sequence Number20136468
Product Code LXH
UDI-Device Identifier07611996112498
UDI-Public07611996112498
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K110135
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number75023360
Device Catalogue Number75023360
Device Lot NumberA77983
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/18/2022
Supplement Date Received by Manufacturer02/15/2022
04/05/2022
Initial Report FDA Received Date02/07/2022
Supplement Report FDA Received Date02/16/2022
04/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/20/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-