| Model Number |
ORT200 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/08/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During a surgical procedure on (b)(6) 2019, the ort200 table drifted downward in the trendelenburg function while the patient was on the table.There were no injuries reported due to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
An imris customer service engineer (cse) was on-site to resolve the issue.The cse flushed the hydraulic oil out of the ort200 table's hydraulic system and replaced with new oil.Afterwards, the cse performed weighted and un-weighted drift testing on the table to verify proper function.
|
| |
|