| Model Number |
227 |
| Medical Device Problem Code |
Suction Problem (2170)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Avanos medical inc.Received a single report that referenced two different incidences, which involved two separate units.This is the first of two reports.Refer to 8030647-2022-00028 for the second report.It was reported that the therapist attempted to suction the patient using a new catheter and the suction did not work.There was no report of injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The product involved in the report has been returned and is being processed for evaluation.A review of the device history record is in-progress.All information reasonably known as of 03 feb 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record for lot 30145570 was reviewed and the product was produced according to product specifications.The actual sample from the reported event was returned for evaluation.Visual examination of the device revealed no breakages or damage.The device was evaluated.Testing of the sample revealed that suctioning occurred slowly when the thumb valve was pressed and suctioning did not occur when the valve was placed in the locked position.The event could not be confirmed as reported; the root cause is unknown.All information reasonably known as of 06 apr 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
| |
|
Search Alerts/Recalls
|