Model Number 429524
Medical Device Problem Code Premature End-of-Life Indicator (1480)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/31/2022
Type of Reportable Event Malfunction
Event or Problem Description
After an implantation period of approx.16 months, it was reported that the device shows the eos status.The icd was explanted.
 
Additional Manufacturer Narrative
Upon receipt, the device interrogation revealed the eos battery status.A number of six charging cycles was documented.The inspection of the icd memory content confirmed a premature detection of the eos battery status on january 31st, 2022.The current consumption of the device in the eos state was tested and was found to be normal.Therefore, the icd was opened to examine the inner assembly.A visual inspection showed no anomalies.The battery voltage was measured revealing a depleted battery.In a next step, the electronic module was attached to an external power supply.The eos status was removed with a technical programmer and the ability of the device to deliver therapies was verified.The anti-bradycardia pacing pulses proved to be normal and in amplitude and frequency as programmed.A fibrillation signal was applied, and the device delivered a defibrillation shock as specified, documenting a normal and expected sensing and shock delivery.In particular, the specified energy level was reached.Further extensive analysis of the electronic components and the device memory content was performed, indicating an intermittent elevated current consumption, that could be narrowed down to an integrated circuit on the electronic module.Based on all available information, the root cause for the intermittent elevated current consumption was, however, not conclusively determinable.The manufacturing records and quality documents for this device were re-investigated.No anomalies were documented during the device manufacturing, in particular during the final acceptance test no abnormality in the current consumption of the device was observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTICOR 7 DR-T DF4 PROMRI
Common Device Name
ICD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key13477014
Report Number1028232-2022-00669
Device Sequence Number5101087
Product Code MRM
UDI-Device Identifier04035479156695
UDI-Public04035479156695
Combination Product (Y/N)N
PMA/510(K) Number
P050023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2022
Device Model Number429524
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/04/2022
Supplement Date Received by Manufacturer05/23/2022
Initial Report FDA Received Date02/07/2022
Supplement Report FDA Received Date05/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/29/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient SexUnknown
-
-