• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET HELICAL NANO 1MM X 3CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET HELICAL NANO 1MM X 3CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0035431030
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/02/2021
Type of Reportable Event Malfunction
Event or Problem Description
The subject device was returned for analysis, and it was discovered that the coil (subject device) was found to be prematurely detached/separated during use.There were no clinical consequences to the patient reported as a result of this event.
 
Additional Manufacturer Narrative
Due to the automated mes (manufacturing execution system) system there are controls in the manufacturing process to ensure the product met specifications upon release.During visual inspection, the main coil was found to be detached and stuck at the tip of the sl10, the detachment seems to be electrical.The main coil was found to be stretched, the suture was found to be damaged, the main coil was found to be severely kinked/bent at the detachment zone, and the coil delivery wire was not returned.During functional inspection, coil jammed ¿ confirmed during the visual inspection, main coil failed/unable to detach ¿ the coil was found to be detached, and main coil false detachment signal ¿ the coil was found to be detached.An attempt was taken to remove the coil from the microcatheter.The device was flushed, and patency mandrel was advanced; however, the coil could not be pushed out.The coil was pulled out of the distal end of the microcatheter for further analysis.The reported complaint was confirmed based on analysis.The device failed to meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The main coil was found to be stretched and kinked/bent.The main coil was found to be detached and stuck at the tip.An assignable cause of procedural factors will be assigned to reported/analyzed code 'coil jammed', and to the analyzed codes 'main coil kinked/bent', 'main coil suture damaged', 'main coil prematurely detached/separated during use' and 'main coil stretched' and also to the reported code 'main coil failed/unable to detach' and 'main coil false detachment signal' as this defects appear to be associated with a product that met stryker and design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural or anatomical factors during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET HELICAL NANO 1MM X 3CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key13477106
Report Number3008881809-2022-00052
Device Sequence Number5437470
Product Code HCG
UDI-Device Identifier04546540698018
UDI-Public04546540698018
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/13/2023
Device Model NumberM0035431030
Device Catalogue NumberM0035431030
Device Lot Number22368491
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/17/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/02/2022
Initial Report FDA Received Date02/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/14/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EXCELSIOR SL-10 (STRYKER MANUFACTURER)
Patient SexUnknown
Patient RaceAsian
-
-